摘要:

根据国内外相关检测标准,确定视觉分辨率和调制传递函数模量作为评价指标,设计检测装置和标靶,编写Matlab程序对采集到的标靶图像进行分析,对13支不同品牌、种类的硬性内窥镜进行检测。实现基于机器视觉检测医用硬性内窥镜分辨率的设备和方法,13支内镜的视觉分辨率为5~16 LP/mm,MTF阈值集中于0.2~0.3之间。结果表明:1)硬性内窥镜镜体、中心、边缘的MTF值均呈现随着线对数增大而减小的趋势;2)每支硬性内窥镜均具有合适的亮度使其分辨率最高;3)硬性内窥镜整体的MTF值偏向由中心决定;4)同一区域下,不同亮度所对应的MTF值相关程度较高,说明内窥镜的分辨率由自身决定。


Research on evaluation of medical rigid endoscope's resolution based on MTF

JI Jun1, JIANG Chang-song2, HOU Chun-wei3, WANG Yun-long1,4, HUANG Zeng-yue1,4, LIU Xin-yu3, XIAO Hong1, YUAN Qing1,4

1. Department of Biomedical Engineering, PLA 305 Hospital, Beijing 100017, China;
2. Department of Biomedical Engineering, Peking University Third Hospital, Beijing 100191, China;
3. Department of Biomedical Engineering, The Fourth Military Medical University, Xi'an 710032, China;
4. Graduate School, Southern Medical University, Guangzhou 510515, China

Abstract: In order to evaluate the resolution of medical rigid endoscopes quantitatively, this article confirmed the visual resolution and modulation transfer function modulus-MTF as the evaluation indexes, designed the testing of target, device and compile Matlab program for automatic analysis and processing of the collected target image to detect 13 different brands, types of rigid endoscopes according to the relevant domestic and international testing standards. The methods and detection equipment based on machine vision were achieved. The visual resolution of 13 endoscopes was ranged from 5 to 16 LP/mm and MTF threshold was between 0.2 and 0.3. The results showed:1)The overall,center,and edge of rigid endoscope's MTF curve showed a trend that MTF values are reducing with logarithmic increasing;2)Rigid endoscope has a appropriate brightness to make the highest resolution; 3)The overall MTF values of rigid endoscope are decided by the center much more;4)in the range of the same region MTF values corresponding to different brightness values have high correlation coefficients. High values of the correlation coefficients reflect the resolution of the endoscope is determined by itself and has little to do with brightness.

Keywords: medical rigid endoscope; resolution; MTF

2014, 40(6): 14-16  收稿日期: 2014-1-24;收到修改稿日期: 2014-3-25

基金项目: 全军军事医学计量科研专项课题(2011-JL2-027);解放军305医院科研基金重点项目(12YA2)

作者简介: 姬军(1970-),男,北京市人,高级工程师,硕士生导师,研究方向为医疗设备计量与质量控制检测技术。

参考文献

[1] 蒋昌松,姬军,黄增跃,等. 医用硬性内窥镜光学性能质控检测技术研究[J]. 中国测试,2013(2):24-26.
[2] Aabakken L,Enochsson L. Mechanics of auality assurance-now and in the future[J]. Best Practice & Research Clinical Gastroenterology,2011(25):419-425.
[3] 吕庆友,官丽梅. 医用内窥镜使用与维修技术[M]. 沈阳:辽宁科学技术出版社,2006:52-150.
[4] GB 11244-2005医用内窥镜及附件通用要求[S]. 北京:中国标准出版社,2005.
[5] SNT 1095-2002进口医用内窥镜性能检验规程[S]. 北京:中国标准出版社,2002.
[6] YY 0068.1-2008医用内窥镜硬性内窥镜第1部分:光学性能及测试方法[S]. 北京:中国标准出版社,2008.
[7] 李林,林家明,王平,等. 工程光学[M]. 北京:北京理工大学出版社,2007:1-23.
[8] 叶春芳. 光学系统MTF快速测量仪的研究[D]. 杭州:浙江大学,2009.
[9] 郭晓兵. 基于图像处理的MTF测试系统研究[D]. 南京:南京理工大学,2009.