基于非介入式X射线管电压测量的研究

2710    2016-01-23

免费

全文售价

作者:杨建1, 侯新生1, 郑永明2, 黄平2

作者单位:1. 成都理工大学, 四川成都 610059;
2. 中国测试技术研究院, 四川成都 610021


关键词:X射线诊断机; 质量控制; X射线管; 非介入式; 管电压


摘要:

为了实现X射线诊断机的质量控制,根据辐射剂量测量的原理,设计了非介入式X射线管电压测量系统。该测量系统以单片机控制系统为平台,智能判断X射线诊断机的照射,通过不同传输路径的射线强度的对比,计算X射线管电压,实现X射线管电压波形的实时存储。实验数据表明,该方法简单有效,安全可靠,适用于X射线诊断机的质量控制,具有广阔的应用前景。


Study on voltage measurement of X-ray tubes based on non-intrusive

YANG Jian1, HOU Xin-sheng1, ZHENG Yong-ming2, HUANG Ping2

1. Chengdu University of Technology, Chengdu 610059, China;
2. National Institute of Measurement and Testing Technology, Chengdu 610021, China

Abstract: In order to control the quality of X-ray diagnostic machine, non-intrusive X-ray tube voltage measurement system was designed according to the theory of radiation dosage testing. This measurement system took the single chip control system as the platform, judged the exposure of X-ray diagnostic machine intelligently, calculated the X-ray tube voltage through the strength contrast of different rays from different transmission path, and stored the real-time wave of the tube voltage simultaneously. The experimental data shows that this method is simple and effective, safe and reliable, applicable to the quality control of the X-ray diagnosis machine. Thus this method has broad application prospects.

Keywords: X-ray diagnostic machine; Quality control; X-ray tube; Non-intrusive; Tube voltage

2008, 34(5): 110-112  收稿日期: 2008-4-22;收到修改稿日期: 2008-7-5

基金项目: 

作者简介: 杨建(1982-),男,江苏南通市人,硕士研究生,专业方向为仪器与测控技术。

参考文献

[1] 魏志勇.医用核辐射物理学[M].苏州:苏州大学出版社, 2003.
[2] 王瑞玉.医用数字乳腺X射线机[M].北京:中国医药科技出版社, 2007.
[3] 黎式棠.医用X射线机原理与维修[M].南宁:广西科学技术出版社, 1999.
[4] 中国标准汇编.医疗器械标准汇编医用X射线卷[S].北京:中国标准出版社, 2002.
[5] 杨勇, 张圆月, 刘志宏.医用诊断X射线剂量探测技术概述[J].中国测试技术, 2005,31(4):94-96.
[6] 孙丽娜, 原培新.X射线安检设备中探测技术研究[J].中国测试技术, 2006,32(3):20-22.
[7] Ranallo F N. Proposed definitions of tube potential(kV) and peak tube potential (kVp) in the non-invasive measurement of X-ray tube potential.Madison, Wisconsin:University of Wisconsin-Madison, 1993:70-80.
[8] International Electrotechnical Commission.Evaluation and routine testing in medical imaging departments-Part 3-2, acceptance tests-Imaging performance of mammographic X-ray equipment[S].IEC Standard 1223-3-2, 1996.